Measurement parameter and Range |
|
Inductance |
|
|
Capacitance |
|
|
Resistance |
R |
0.01m ~199.99M |
|
|
Impedance |
|Z| |
0.01m ~199.99M |
|
|
Admittance |
|Y| |
5.00nS (199.99M )
~100.00S (100m ) |
|
|
Dissipation factor |
|
|
Quality |
|
|
Equivalent series resistance |
RS |
0.01m ~199.99M |
|
|
Equivalent parallel resistance |
RP |
0.01m ~199.99M |
|
|
Conductance |
G |
5.000nS~?i199.99M ?j
?`100.00S (100?? ) |
|
|
Reactance |
X |
0.01m ~199.99M |
|
|
Susceptance |
B |
5.000nS (199.,99M )
~100.00S (100m ) |
|
|
Phase Angle |
|
|
Measurement terminals |
Coaxial BNC connectors: FORCE (2), SENSE (2)
Binding post: GUARD (1) |
|
basic accuracy |
0.1% |
|
Measurement frequency |
42Hz~5.00MHz (±0.01%) |
|
Measrement signal level |
0.01~5.00Vrms/0.01~99.99mA rms
(42.0Hz~1.00MHz)
0.05~1.00Vrms/0.05~20.00mA rms
(1.01~5.00MHz) |
|
Signal source impedance |
50 |
|
Measurement speed |
18msMIN (at 10kHz)
20kHz~5.00MHz 40ms max
42Hz~19.0kHz 40ms ~ |
|
|
Measurement circuit mode |
Series equivalent or Parallel equivalent. |
Zero correction |
You can automatically correct parasitic impedance with the circuit open or short-circuited for zero balancing. Zero correction in open mode is executable with 1k or higher and that in short mode, with less than 1k |
Memory |
9 testing conditions storable in memory |
GO/NO Judgment |
High/low limits can be set in a range of 00000 to ±19999 and each subject device can be classified into 10 sorts according to the judgment (absolute value or percentage). |
Display |
DISP-A/DISP-B
4-1/2digit display |
Judgment results output |
DISP-A/DISP-B LO, HI, GO, TOTAL GO, TOTAL NO-GO; BUZZER ON (Isolated) |
Output signals |
MEASURE END, TEST BUSY, BIN, ERROR (Isolated) |
Guaranteed temperature range |
23°C ±5°C (Specified accuracies are guaranteed in this temperature range. In a range of 5°C to 40°C, limits of accuracy are doubled.) |
External control signals |
CONTROL PANEL/EXT, EXT TRIGGER, PANEL No (Isolated) |
Signals |
RS-232C standard provision |
Operating temperature range |
0~50°C |
Power requirements |
100/120/220/240V AC ±10%, 50/60Hz; approx. 60VA |
Dimensions & Mass |
250(W) x 148(H) x 400(D) mm, approx. 7kg |
Options |
Test adapter (KCA-06, KCA-07,KCA-08, KCA-09N),
BCD Board(KCB-51),